MC1200 High Accuracy
Dual Display Multifunction Temperature and Process
Calibrator
with NIST
Calibration Certificate
Precision Dual Display Multifunction Calibrator
|
| MC1200 Multifunction Process Calibrator |

|
| Supplied with Protective Rubber Boot |

| MC1200 Large Easy to Read Display |

|
|
The MC1200 multifunction
calibrator provides a feature set unmatched in high accuracy, handheld calibrators in its price range. The MC1200 calibrator
provides the functions and accuracy associated with
fixed-installation, laboratory instrument, and has everything needed for virtually any calibration task. Measure and source
thermocouples, RTDs, current, voltage, and frequency, and source pulse trains.
A
communications port compatible with both Fluke® 700 Series and Beta Calibrators pressure modules is provided, as is an isolated mA/V read-back circuit. Arrow keys, direct numeric keypad entry, and three software driven
function buttons, plus a large backlit, menu-driven graphics display combine to provide a highly intuitive, simple yet powerful
operator interface. With a built-in 250 Ohm resistor for
HART®
compatibility, the MC1200 is compatibility with smart transmitters and PLCs,
full fuse-less protection, are just some of the additional features that make the MC-1200 the single most indispensable tool available for virtually any calibration task.
Features
•
Measure and source Thermocouples (13
types), RTDs (13 types), Ohms,
current, voltage, frequency
and source pulse trains signals
•
Isolated mA/V read-back circuit for
complete transmitter calibration
• Built-in 250 Ohm resistor for HART® compatibility • Communication port compatible with both Fluke 700 Series and
Beta Calibrator pressure modules
• Built-in 24V supply can drive
4 to 20mA loops up to 1000 Ohms •
Direct entry of custom RTD
coefficients (Ro, A, B,
C) • All source modes can be
programmed with dedicated setpoints
to speed calibration and linearity
tests • Highest accuracy in class to
0.015% of reading • Meets CE
requirements and is designed to IEC
1010 safety standards
Click here to view a Flash Demonstration of the MC1200
Specifications
Environmental
Operating Temperature: -10°C to +50°C
(-14°F to 122°F) Storage Temperature: -20°C to +70°C (-4°F to 158°F)
Stability:
±0.005% of reading/°C outside of 23 °C ± 5°C
Power Requirements: 6Vdc; Batteries
4 AA alkaline (supplied) or optional rechargeable
Mechanical Dimensions: 8.7" H x 4.2" W x 2.3" D (220.9 x 106.6 x 58.4 mm)
Weight: 30.5
ounces (863grams)
|
Signal Specifications Voltage
Read and Source Source:
0.000 to 20.000Vdc Read: Isolated 0.000 to 30.000Vdc
Non-Isolated 0.000 to 20.000Vdc Thermocouple Types and Ranges: See
table below Accuracy: See Table Below
Temperature scale: ITS90 mV Read/Source:
-10.000 to +75.000 mV Current (mA) Source:
0.000 to 24.000mA Read: Isolated: 0.000 to 24.000mA; Non-Isolated: 0.000 to 24.000mA
|
Frequency (1 to 20 V selectable amplitude) CPM Source and Read: 2.0
to 600.0 CPM Hz Source and Read: 1.0 to 1000.0 Hz kHz Source and Read: 1.0 to 10.00 kHz Pulse (Source only; 1 to 20 V selectable amplitude) Pulses: 1 to 30,000.0 - 2 CPM to
10 kHz Ohms Source: 5.0 to 4000 Ohms Read: 0.00 to 4000.0
Ohms
|
Thermocouple Range and Accuracy
| TC Type | Range °C (°F) | Accuracy* | | | -210.0 to 0.0°C (-346 to
32°F) | 0.4°C
(0.7°F) |
|
J | 0 to
800.0°C (32 to 1472°F) | 0.2°C (0.4°F) |
|
| 800.0 to 1200.0°C (1472 to 2192°F) | 0.3°C (0.5°F) |
|
| -200.0 to 0.0°C (-328 to 32°F) | 0.6°C (1.1°F) |
|
K | 0 to 1000°C (32 to 1832°F) | 0.3°C (0.5°F) |
|
| 1000.0 to 1372.0°C (1832 to 2502°F) | 0.5°C (0.9°F) |
|
T | -250.0 to 0.0°C (-418 to 32°F) | 0.6°C (1.1°F) |
|
| 0 to 400.0°C (32 to 752°F) | 0.2°C (0.4°F) |
|
E | -250.0 to -100.0°C (-418 to 212°F) | 0.6°C (1.1°F) |
|
| -100.0 to 1000.0°C (-148 to 1832°F) | 0.2°C (0.4°F) |
|
R | 0 to 1767.0°C (32 to 3212°F) | 1.2°C (2.2°F) |
|
S | 0 to 1767.0°C (32 to 3212°F) | 1.2°C (2.2°F) |
|
| 600.0 to 800.0°C (1112 to 1472°F) | 1.2°C (2.2°F) |
|
B | 800.0 to 1000.0°C (1472 to 1832°F) | 1.3°C (2.4°F) |
|
| 1000.0 to 1820.0°C (1832 to 3308°F) | 1.5°C (2.7°F) |
|
C | 0 to 1000.0°C (32 to 1832°F) | 0.6°C (1.1°F) |
|
| 1000.0 to 2316.0°C (1832 to 3320°F) | 2.3°C (4.2°F) |
|
XK | -200.0 to 800.0°C (-328 to 1472°F) | 0.2°C (0.4°F) |
|
BP | 0 to 800.0°C (32 to 1472°F) | 0.9°C (1.6°F) |
|
| 800.0 to 2500.0°C (1472 to 4532°F) | 2.3°C (4.2°F) |
|
L | -200.0 to 0.0°C (-328 to 32°F) | 0.25°C (0.45°F) |
|
| 0 to 900.0°C (32 to 1652°F) | 0.2°C (0.4°F) |
|
U | -200.0 to 0.0°C (-328 to 32°F) | 0.5°C (0.9°F) |
|
| 0 to 600.0°C (32 to 1112°F) | 0.25°C (0.45°F) |
|
N | -200.0 to 0.0°C (-328 to 32°F) | 0.8°C (1.4°F) | | | 0 to 1300.0°C (32 to 2370°F) | 0.4°C (0.7°F) |
*All TC errors include CJC errors
CJC error outside of 23 ±
5°C is 0.05°C/°C (In °C add .2 for cold junction
compensation error.)
|
RTD Type |
Range |
Accuracy* |
|
|
-200.0 to 100.0°C (-328 to 212°F) |
0.2°C (0.4°F) |
|
Pt100 (385) |
100.0 to 300.0°C (212 to 572°F) |
0.3°C (0.5°F) |
|
|
300.0 to 600.0°C (572 to 1112°F) |
0.4°C (0.7°F) |
|
|
600.0 to 800.0°C (1112 to 1472°F) |
0.5°C (0.9°F) |
|
|
-200.0 to 100.0°C (-328 to 212°F) |
0.8°C (1.4°F) |
|
Pt200 (385) |
100.0 to 300.0°C (212 to 572°F) |
0.9°C (1.9°F) |
|
|
300.0 to 630.0°C (572 to 1166°F) |
1.0°C (1.8°F) |
|
|
-200.0 to 100.0°C (-328 to 212°F) |
0.4°C (0.7°F) |
|
Pt500 (385) |
100.0 to 300.0°C (212 to 572°F) |
0.5°C (0.9°F) |
|
|
300.0 to 630.0°C (572 to 1166°F) |
0.6°C (1.1°F) |
|
|
-200.0 to 100.0°C (-328 to 212°F) |
0.2°C (0.4°F) |
|
Pt1000 (385) |
100.0 to 300.0°C (212 to 572°F) |
0.3°C (0.5°F) |
|
|
300.0 to 630.0°C (572 to 1166°F) |
0.4°C (0.7°F) |
|
|
-200.0 to 100.0°C (-328 to 212°F) |
1.4°C (2.5°F) |
|
Pt385-10 |
100.0 to 300.0°C (212 to 572°F) |
1.6°C (2.9°F) |
|
|
300.0 to 600.0°C (572 to 1112°F) |
1.8°C (3.2°F) |
|
|
600.0 to 800.0°C (1112 to 1472°F) |
2.0°C (3.6°F) |
|
|
-200.0 to 100.0°C (-328 to 212°F) |
0.4°C (0.7°F) |
|
Pt385-50 |
100.0 to 300.0°C (212 to 572°F) |
0.5°C (0.9°F) |
|
|
300.0 to 600.0°C (572 to 1112°F) |
0.6°C (1.1°F) |
|
|
600.0 to 800.0°C (1112 to 1472°F) |
0.7°C (1.3°F) |
|
|
-200.0 to 100.0°C (-328 to 212°F) |
0.2°C (0.4°F) |
|
Pt100 (3926) |
100.0 to 300.0°C (212 to 572°F) |
0.3°C (0.5°F) |
|
|
300.0 to 630.0°C (572 to 1166°F) |
0.4°C (0.7°F) |
|
|
-200.0 to 100.0°C (-328 to 212°F) |
0.2°C (0.4°F) |
|
Pt100 (3916) |
100.0 to 300.0°C (212 to 572°F) |
0.3°C (0.5°F) |
|
|
300.0 to 630.0°C (572 to 1166°F) |
0.4°C (0.7°F) |
|
Ni120 (672) |
-80.0 to 260.0°C (-112 to 500°F) |
0.2°C (0.4°F) |
|
Cu10 |
-100.0 to 260.0°C (-148 to 500°F) |
1.4°C (2.5°F) |
|
Cu50 |
-180.0 to 200.0°C (-292 to 392°F) |
0.4°C (0.7°F) |
|
Cu100 |
-180.0 to 200.0°C (-292 to 392°F) |
0.3°C (0.5°F) |
|
YSI400 |
15.00 to 50.00 (59 to 122°F) |
0.1°C (0.2°F) |
*Read
Accuracy is based on 4-wire input. For 3-wire input add ±0.05Ω
assuming all three RTD leads are matched.
|
Process Signals |
Read Accuracy* |
Source Accuracy* |
|
Current (mA) |
±0.015% of reading,±2 μA |
±0.015%
of reading, ±2 μA |
|
CPM |
±0.05% of reading,±1 LSD |
±0.05%
of reading |
|
Hz |
±0.05% of reading,±1 LSD |
±0.05%
of reading |
|
kHz |
±0.05% of reading, ±1 LSD |
±0.125%
of reading |
|
Ohms works with all pulsed transmitters down to 5 ms |
|
|
|
400 Ohm Range |
±0.025% of reading, ±0.05W |
±0.025% of reading, ±0.05W |
|
4000 Ohm Range |
±0.025% of reading, ±0.5W |
±0.025%
of reading, ±0.5W |
*MC-1200 Accuracy Specifications (23 °C ±5 °C unless
otherwise noted)
|
 |
|
|
|
Each instrument includes Instruction Manual, Test Leads, Protective
Rubber Boot and NIST Certificate.
|
|
 |
|
|
|
|
|
Multifunction Temperature and Process Calibrators, Thermocouple and mA Calibrators, Thermocouple and RTD Dry Block Calibrators, Portable Infrared Thermometers, Fixed Mounted Industrial Infrared Thermometers, Thermocouple and RTD 2-wire Head Mounted Transmitters, DIN Rail Mounted Thermocouple and RTD Transmitters, Temperature and Process Panel Meters, Temperature & Humidity Transmitters, Hand Held Digital Thermocouple Thermometers, MIST Microprocessor Based Integral Transmitter RTD Temperature
Sensors,
Industrial Test tools and Maintenance Kit, Indoor Air
Quality (IAQ) and Test Meters, Hand Held Humidity and Air Flow Meters, Industrial Portable
Refrigerant and Gas Leak Detection Meters, Pressure Measurement and Calibration Instruments
Temperature & Process Instruments Inc. * 1767 Central Park Avenue * Suite 112 * Yonkers